Intex Test Procedure
Main
Intex Allergy Lab
Procedure
Ordering

INTEX Allergen Specific IgE-Test:
BRIEF PROCEDURE DESCRIPTION

The package comes with a full manual, however here is also a brief outline of the procedure involved.

Any comments or questions? Please contact us at

info@allergy-lab.com

Intex is supplied by OLRON SYSTEMS Ltd.

1. Preparatory Steps

  • Bring all reagents and sample to room temperature (RT)
  • Tear open blister card and remove strip
  • Cut blister as shown
  • Mark strip and fill in the patient card

 

2.  First immunological reaction

  • Fill blister no. 1 to the marking (1) with serum or heparinized plasma
  • Place strip into blister
  • remove air bubbles by vertical movements of the dipstick
  • Incubate overnight (16-24 h) at RT

 

3. First wash

  • Incubate strip in distilled water at RT for 10 minutes

 

  • Remove water by gently blotting with a paper towel

 

4. Second immunological reaction

 

  • Fill blister no.2 to the marking (1) with Conjugate IgE sp.
  • Place strip into blister
  • Remove air bubbles by vertical movements of the strip
  • Incubate 1hr at RT

5. Second Wash

  • Incubate dipstick in distilled water 2 X 10 minutes at RT

 

  • Remove water by gently blotting with a paper towel

6. Enzyme Reaction

  • Fill blister no. 3 to the first marking (1/6) with Chromagen.
  • Add a Substrate Buffer to the second marking
  • Place strip into blister and mix solutions by vertical movements of the dipstick

 

7. Third Wash

  • Incubate strip in distilled water for 5 minutes at RT

  • Remove water by gently blotting with a paper towel
  • Dry strip in the dark for 30 minutes at RT

 

8. Evaluation of Resluts by Color Card or Densitometer

  • Determine INTEX Class by comparison with the color card, or use the Densitometer
  • Introduce results into a patient card.
[Main] [Intex Allergy Lab] [Procedure] [Ordering]

INTEX is supplied by OLRON SYSTEMS Ltd. c/o Turner Clinic, Old Dublin Road, Stillorgan, Blackrock, Co.Dublin, Ireland. Phone +353 1 2783806.